From Anvita Health Wiki
Indications
-
Reference-interval
-
Principle
- The Kodak Ektachem Clinical Chemistry Slide (Mg) is a dry, multilayered analytical element coated on a clear polyester support.
- A 10 microliter drop of sample is deposited on the slide & evenly distributed by the spreading layer into the underlying layers. Solute molecules penetrate the scavenger layer where calcium is chelated from the sample. Magnesium (both free & protein-bound) from the sample then reacts with the formazan dye derivative in the reagent layer; the high magnesium affinity of the dye dissociates magnesium from binding proteins. The resulting magnesium-dye complex causes a shift in the dye absorption maximum from 540 nm to 630 nm. The amount of dye complex formed is proportional to the magnesium concentration present in the sample and is measured by reflection density.
Clinical-significance
Increases
-
Decreases
-
Specimen
More General Terms
Additional Terms
References
- Kodak Ektachem 700 Test Methodologies Manual, Kodak Clinical Products, Rochester, N.Y., 1990.
- Kodak Ektachem 700 Operators Manual, Kodak Clinical Products, Rochester, N.Y., 1987.
- Package Insert, Kodak Ektachem Special Calibrators, Kodak Clinical Products, Rochester, N.Y., 1985.
- Package Insert, Bio-Rad Liquichek Controls, Bio-Rad ECS Division, Anaheim, CA., 1992.
magnesium (Mg+2) in serum